Test Name | Procedure |
RPM | Insert test bur, attach to motor, turn on cooling air turn on water, turn motor on & gradually increase speed to max. (40K / 200K on Suretest-E) |
Current draw @ 60k to 200k RPM | Turn motor on to 40K / 200K, run the motor(alone) and record the amperage reading from the gage on the unit. Attach handpiece and repeat, recording the second reading. Subtract the first reading from the second, this equals the handpiece current consumption |
Free run time | Run for 30 to 45 sec. w/no load, then check for any heat build up on and around the head & neck areas |
Cooling air & water | Required to properly test handpiece |
Audible test free run & under load | Check for vibration and running noise after radial & axial loading for both clockwise and counterclockwise operation. When loading axially, the clamp collar must not touch the press button. Touch to inspect for heat. |
Test Name | KaVo 25LHA/25LPA/25LCA/25LN/ M05/M25/E25C/E25L | Bien Air CA1:4-43 1:4/CA1541 1:5 |
RPM | 200K | 160K/200K |
Current draw @ 60k to 200k RPM | Not to exceed 0.150amp | Not to exceed 0.150amp |
Free run time | No evidence of heat should exist. | Little to no heat should exist. |
Cooling air & water | On | On |
Audible test free run & under load | Check for vibration and running noise after radial & axial loading for both clockwise and counterclockwise operation. When loading axially, the clamp collar must not touch the press button. Touch to inspect for heat. | Check for vibration and running noise after radial & axial loading for both clockwise and counterclockwise operation. When loading axially, the clamp collar must not touch the press button. Touch to inspect for heat. |
Test Name | MW eStylus 1:5 / E Electric | Sirona C200L/C200/BA250T/BAS250LT/MAXIMA ELITE 1:5 |
RPM | 200K | 200K |
Current draw @ 60k to 200k RPM | Not to exceed 0.150amp / 0.125amp | Not to exceed 0.125amp |
Free run time | Little to no heat should exist | No evidence of heat should exist |
Cooling air & water | On | On |
Audible test free run & under load | Check for vibration and running noise after radial & axial loading for both clockwise and counterclockwise operation. When loading axially, the clamp collar must not touch the press button. Touch to inspect for heat. | Check for vibration and running noise after radial & axial loading for both clockwise and counterclockwise operation. When loading axially, the clamp collar must not touch the press button. Touch to inspect for heat. |
Test Name | Star Titan E-lectric 1:5 / NuTorque | NSK Ti-MAX Ti95L/Ti95/X95L/M95L/Z95L / Brasseler Forza F5 |
RPM | 200K | 200K |
Current draw @ 60k to 200k RPM | Not to exceed 0.135amp | Not to exceed 0.135amp |
Free run time | No evidence of heat should exist. | No evidence of heat should exist |
Cooling air & water | On | On |
Audible test free run & under load | Check for vibration and running noise after radial & axial loading for both clockwise and counterclockwise operation. When loading axially, the clamp collar must not touch the press button. Touch to inspect for heat. | Check for vibration and running noise after radial & axial loading for both clockwise and counterclockwise operation. When loading axially, the clamp collar must not touch the press button. Touch to inspect for heat. |
Test Name | Star Titan E-lectric 1:5 / NuTorque |
RPM | 200K |
Current draw @ 60k to 200k RPM | Not to exceed 0.125amp |
Free run time | No evidence of heat should exist. |
Cooling air & water | On |
Audible test free run & under load | Check for vibration and running noise after radial & axial loading for both clockwise and counterclockwise operation. When loading axially, the clamp collar must not touch the press button. Touch to inspect for heat. |